Determination of total dissolved phosphorus in water samples by axial inductively coupled plasma atomic emission spectrometry
Tipo de material:
TextoSeries ; Fresenius J Anal Chem, 360(1), p.132-136, 1998Trabajos contenidos: - De Boer, J.L.M
- Kohlmeyer, U
- Breugem, P.M
- Van Der Velde-Koerts, T
| Item type | Current library | Collection | Call number | Status | Date due | Barcode | |
|---|---|---|---|---|---|---|---|
Documentos solicitados
|
CICY Documento préstamo interbibliotecario | Ref1 | B-8780 (Browse shelf(Opens below)) | Available |
Browsing CICY shelves, Shelving location: Documento préstamo interbibliotecario, Collection: Ref1 Close shelf browser (Hides shelf browser)
A procedure was developed to determine total dissolved phosphorus in groundwater-like samples using axial ICP-AES. Severe and peculiar matrix effects (strongly positive at lower and leveling off at higher matrix concentrations)in the presence of Na and Ca were observed. To reduce these matrix effects, a double approach was utilized consisting of a so-called minimum matrix in combination with an internal standard (Ga in this case). The 'minimum matrix' (small amounts of K, Mg and Na)was only added to the standard solution(s)used for the calibration. The detection limit for the whole procedure was 12 mg/L using the P213 nm line. Residual matrix effects were less than 3