TY - BOOK AU - Salvo,P. AU - Calisi,N. AU - Melai,B. AU - Cortigiani,B. AU - Mannini,M. AU - Caneschi,A. AU - Di Francesco,F. TI - Temperature and pH sensors based on graphenic materials KW - PH KW - TEMPERATURE KW - GRAPHENE KW - FLEXIBLE SENSOR N2 - Point-of-care applications and patients' real-time monitoring outside a clinical setting would require disposable and durable sensors to provide better therapies and quality of life for patients. This paper describes the fabrication and performances of a temperature and a pH sensor on a biocompatible and wearable board for healthcare applications. The temperature sensor was based on a reduced graphene oxide (rGO)layer that changed its electrical resistivity with the temperature. When tested in a human serum sample between 25 and 43 °C, the sensor had a sensitivity of 110±10 ?/°C and an error of 0.4±0.1 °C compared with the reference value set in a thermostatic bath. The pH sensor, based on a graphene oxide (GO)sensitive layer, had a sensitivity of 40±4 mV/pH in the pH range between 4 and 10. Five sensor prototypes were tested in a human serum sample over one week and the maximum deviation of the average response from reference values obtained by a glass electrode was 0.2 pH units. For biological applications, the temperature and pH sensors were successfully tested for in vitro cytotoxicity with human fibroblast cells (MRC-5)over 24 h UR - https://drive.google.com/open?id=1iLkF2FZ_bNgSa24_It6R-EexhVlZ0nVr&usp=drive_copy ER -