245 1 0 _aA Scanning Michelson Interferometer for the Measurement of the Concentration and Temperature Derivative of the Refractive Index of Liquids
490 0 _vBerichte der Bunsengesellschaft für physikalische Chemie, 99(4), p.600-608, 1995
520 3 _aA compact scanning Michelson interferometer for the measurement of both the concentration and temperature derivative of the refractive index of liquids, liquid mixtures, and solutions has been developed. Scanning of one of the mirrors allows the extraction of the phase information and makes the instrument insensitive to intensity drifts of the light source. The refractive index noise level of the empty instrument is of the order of 10?8 for a 10 mm sample cell, and for continuously varying refractive indices there is no upper limit of the linear detection range. Accuracies of the order of 0.5 per cent can be achieved both for (n/c)p,T and (n/T)p,c.
650 1 4 _aLIQUIDS
650 1 4 _aREFRACTIVE INDEX
700 1 2 _aBecker, A.
700 1 2 _aKohler, W.
700 1 2 _aMuller, B.
856 4 0 _uhttps://drive.google.com/file/d/1QffflXP6zHeOuP582fJSuaAndq1G07HL/view?usp=drivesdk
_zPara ver el documento ingresa a Google con tu cuenta: @cicy.edu.mx
942 _2Loc
_cREF1
008 250602s9999 xx |||||s2 |||| ||und|d
999 _c50169
_d50169