Design and analysis of a burst strength device for testing vascular grafts

Tipo de material: TextoTextoSeries ; Review of Scientific Instruments, 90(1), p.014301, 2019Trabajos contenidos:
  • Pérez-Aranda, C
  • Gamboa, F
  • Castillo-Cruz, O
  • Cauich-Rodríguez, J. V
  • Avilés, F
Tema(s): Recursos en línea: Resumen: The design and analysis of a device to measure the burst strength (strength under a state of pure radial internal pressure)and compliance of vascular grafts and flexible pressurized tubes is presented. The device comprises three main sections, viz., a clean air-dry pressure controller, a test specimen holder, and automated software for control and data collection. Air pressure is controlled by means of a valve and a dedicated mechanism allowing reaching up to 120 psi in increments of 1 psi, and recording pressure changes with 0.04 psi resolution. The circumferential strain is determined by measuring the radial displacement of the vascular graft using an optical arrangement capable of determining a maximum radial displacement of 10 mm with 0.02 mm resolution. The instrument provides a low uncertainty in compliance (±0.32 per cent/100 mm Hg?1)and burst strength measurements. Due to its simplicity, the device can easily be reproduced in other laboratories contributing to a dedicated instrument with high resolution at low cost. The reliability of the apparatus is further confirmed by conducting finite element analysis, elasticity solutions for pressurized cylinders, and testing of small diameter vascular grafts made of a commercial aliphatic polyurethane tested under radial internal pressure.
Tags from this library: No tags from this library for this title. Log in to add tags.
Star ratings
    Average rating: 0.0 (0 votes)

The design and analysis of a device to measure the burst strength (strength under a state of pure radial internal pressure)and compliance of vascular grafts and flexible pressurized tubes is presented. The device comprises three main sections, viz., a clean air-dry pressure controller, a test specimen holder, and automated software for control and data collection. Air pressure is controlled by means of a valve and a dedicated mechanism allowing reaching up to 120 psi in increments of 1 psi, and recording pressure changes with 0.04 psi resolution. The circumferential strain is determined by measuring the radial displacement of the vascular graft using an optical arrangement capable of determining a maximum radial displacement of 10 mm with 0.02 mm resolution. The instrument provides a low uncertainty in compliance (±0.32 per cent/100 mm Hg?1)and burst strength measurements. Due to its simplicity, the device can easily be reproduced in other laboratories contributing to a dedicated instrument with high resolution at low cost. The reliability of the apparatus is further confirmed by conducting finite element analysis, elasticity solutions for pressurized cylinders, and testing of small diameter vascular grafts made of a commercial aliphatic polyurethane tested under radial internal pressure.

There are no comments on this title.

to post a comment.